Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2003;45(4):456-461.
DOI: https://doi.org/10.4097/kjae.2003.45.4.456   
Cardiac Output Estimations by Esophageal Doppler Cannot Replace Estimations by the Thermodilution Method in Off-pump Coronary Artery Bypass Surgery Patients.
Yon Hee Shim, Young Jun Oh, Sang Beom Nam, Jong Hwa Lee, Ho Dong Rhee, Young Lan Kwak
Department of Anesthesiology and Pain Medicine and Anesthesia and Pain Research Institute, Yonsei University College of Medicine, Seoul, Korea. ylkwak@yumc.yonsei.ac.kr
Abstract
BACKGROUND
Esophageal doppler is discribed as a non-invasive alternative to cardiac output (CO) estimation by thermodilution, the current bedside "gold standard". This study was designed to evaluate the accuracy of CO estimations performed by esophageal doppler (EDCO), compared to those obtained using a continuous CO pulmonary flotation catheter (TDCO).
METHODS
In 16 patients undergoing off-pump coronary artery bypass surgery, CO was measured simultaneously by the esophageal doppler and the thermodilution method, after induction (A), after sternotomy (B), after coronary revascularization (C), and after sternal closure (D). Agreement between the TDCO and EDCO estimations was assessed by analyzing their mean differences and the distribution of these differences. Relative CO changes (percentages of the previous value) was analyzed by the same method.
RESULTS
Both absolute CO values and relative CO changes by esophageal doppler showed a considerable scatter compared to those obtained using the thermodilution method. The bias (EDCO-TDCO) between the two mehtods was -0.8 +/- 2.7 L/min for A, -0.9 +/- 2.5 L/min for B, -0.9 +/- 3.6 L/min for C, and -0.6 +/- 2.7 (mean +/- 2 SD) L/min for D. On analyzing changes in CO, no significant method bias was found but 2 SD of the bias were +/- 74% for A to B, +/- 100% for B to C, and +/- 83% for C to D.
CONCLUSIONS
These results suggest that CO estimations by esophageal doppler cannot replace estimations by the thermodilution method in patients undergoing off-pump coronary artery bypass graft surgery.
Key Words: cardiac output; esophageal doppler; monitoring; off-pump coronary artery bypass graft; thermodilution


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
(서울특별시 마포구 마포대로 109 롯데캐슬 프레지던트 101동 3503호)
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                
Business Name: Korean Society of Anesthesiologists (대한마취통증의학회)
Business Registration: 106-82-07194
Representative: Young-Tae Jeon (전영태)

Copyright © 2025 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next