Korean J Anesthesiol Search

CLOSE


Korean Journal of Anesthesiology 2004;46(6):747-750.
DOI: https://doi.org/10.4097/kjae.2004.46.6.747   
A Case of Pulmonary Artery Catheter Malposition after the Weaning of Cardiopulmonary Bypass: A case report.
Jong Hyun Lee, Young Ho Jang
Department of Anesthesiology and Pain Medicine, School of Medicine, Keimyung University, Daegu, Korea.
Abstract
A 79-year old male with angina underwent emergent coronary artery bypass graft. After the induction of anesthesia, a pulmonary artery catheter (PAC) was placed via the right subclavian vein. At weaning from cardiopulmonary bypass, the pulmonary artery wave form did not appear on the monitor. We pulled and pushed the PAC repeatedly. However, we could observed no pulmonary artery pressure; the PAC was not pulled out at the 42 cm from the distal tip. Chest X-ray revealed a malposed PAC into the ipsilateral internal jugular vein. No deleterious complications were observed either during the immediate or later postanesthetic period.
Key Words: complication; monitoring; pulmonary artery catheter