Korean Journal of Anesthesiology 2001;40(1):11-15.
DOI: https://doi.org/10.4097/kjae.2001.40.1.11   
Identification of Epidural Space with a Pressure Detection Apparatus.
Jin Ho Bae, Byung Dal Lee, Sang Bum Kim, Sang tea Kim, Eun Jong Cha, Seung Woon Lim
1Department of Anesthesiology, Graduate School, Chungbuk National University, Cheongju, Korea.
2Department of Anesthesiology, Samsung Medical Center, Seoul, Korea.
3Department of Biomedical Engineering, Chungbuk National University, Cheongju, Korea.
Abstract
BACKGROUND
Techniques for identification of the epidural space have advantages and disadvantages. We made an electronic apparatus to ensure an epidural space by detecting the reduction of pressure. We investgated the instrument to see if this could be adequately used for identification of the epidural space.
METHODS
Thirty adult patients scheduled to receive an epidural injection for surgery were selected for the investigation. We connected the apparatus with an epidural Tuohy needle of which the tip was in the ligamentum flavum. We injected air to increase the pressure inside the epidural needle to 50 mmHg or 100 mmHg, and then the epidural needle was advanced slowly until we heard the alarm from the epidural detector when pressure inside the epidural needle suddenly decreased under the set point, suggesting the epidural needle reached the epidural space. After ensuring the epidural space we injected the anesthetics to induce epidural anesthesia.
RESULTS
In both groups, we obtained successful anesthesia results and there were no complications. The volume of air injected into the epidural space was 0.64 +/- 0.1 in 50 mmHg group and 0.95 +/- 0.2 in 100 mmHg group.
CONCLUSIONS
The use of the epidural detector in 30 patients demonstrated that this instrument could be a safe and effective means of identifying the epidural space. It allows the reduction of air injected into the epidural space, reducing the chance of complications due to an overinjection of air.
Key Words: Anesthetetics techniques: epidural; Equipment: pressure dectection apparatus


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next