Korean Journal of Anesthesiology 1995;28(5):611-617.
DOI: https://doi.org/10.4097/kjae.1995.28.5.611   
Accuracy and Continuity of Infusion Devices with Volumetric Analyzer.
Jung Soo Yoon, Hwang Jung Kim, Jin Yun Kim, Sun Gyoo Park
Department of Anesthesiology, College of Medicine, Chung-Ang University, Seoul, Korea.
Abstract
The proliferation of high-technology infusion devices has greatly influenced the practice of anesthesia by allowing precise delivery intraoperatively and postoperatively of intravenous anesthetics, analgesics, and vasopressors. During continuous drug administration to pediatric patients, unfavorable pharmacologic effects have occured. These effects were attributed to variations in flow from infusion devices. The intent of this investigation was to evaluate the influence of infusion devices on the accuracy and continuity effused fluid. Hartmann's solution was run through eight infusion pump at 5, 10, 15, 20, 25, 30, 40, 60, 80 and 100ml/hr. The volume of the fluid was measured for one hour at each flow rate using a infusion device analyzer(IDA-2) in vitro. Flow accuracy(ml/hr) was calculated by dividing the effused volume by the effusion time ; this value was subsequently expressed as a percentage of the desired flow rate. Flow continuity was expressed by the coefficient of variance(CV); analysis of variance with SAS was used for comparisons. Results were as follows ; 1) The flow accuracy values ranged from 98.8% to 114.2% of the desired rate. 2) Seven infusion pumps had flow rates within 5% of the desired rates at each flow rate but one pump had more than 10% of desired rate. 3) Each of the respective infusion pumps produced various levels of flow continuity and ranged from 0.7% to 2.5%(CV value). Conclusion , further evaluations are needed to differentiate continuity from other infusion devices(e.g., syringe infusion pump) in vitro and evaluations are needed to define the clinical significance in vivo. Future use of flow-rate data will likely be of significant value in establishing policies and procedures for the utilization of infusion devices and guidelines for product selection.
Key Words: Infusion pump; Flow accuracy; Flow continuity
TOOLS
Share :
Facebook Twitter Linked In Line it
METRICS Graph View
  • 3 Crossref
  •    
  • 1,808 View
  • 38 Download


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
(서울특별시 마포구 마포대로 109 롯데캐슬 프레지던트 101동 3503호)
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                
Business Name: Korean Society of Anesthesiologists (대한마취통증의학회)
Business Registration: 106-82-07194
Representative: Young-Tae Jeon (전영태)

Copyright © 2025 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next