Korean Journal of Anesthesiology 1987;20(6):782-787.
DOI: https://doi.org/10.4097/kjae.1987.20.6.782   
Measarement of Drop Factor of IV Infusion Set with Infusion Pump.
Yoon Kee Kim, Kyoung Hun Kim, Dong Ho Lee, Kyo Sang Kim, Jung Kook Suh, Hee Koo Yoo, Ik Sang Seung, Se Ung Chon
Department of Anesthesiology, College of Medicine, Hanyang University, Seoul, Korea.
Abstract
Accurate and reliable administration of intravenous infusion is necessary for a variety of different therapies including constant heparin, parenteral nutrition, cardiotonic and vasoactive mediciation and chemotherapy. For these purposes, drop counting is a common practice and a convenient way to measure the flow rate and the dose rate, but it is an unreliable and unexact method because the intravenous infusion sets from various manufacturers deliver different-sized drops depending on the rate of formation of the drops, properties of the particular fluid and physical characteristics of the intravenous infusion sets. This study evaluated the actual drop factor (DF) of a variety of intravenous infusion sets (10 IV Sets : 5 marked and 5 unmarked DF) at a low flow rate setting (20 ml/hr for pediatric and 40 ml/hr for adult) using a peristaltic fluid pump and distilled water, The measured Df is obtained by dividing the collected fluid volume per hour (ml/hr) by the number of drops per minute (drops/min) on the peristaltic fluid pump setting. The results ware as follows : 1) The actual DF values of the marked lV sets were 2 sets (DF= 1) 0.91 +/-0.04 and 0.97+/-0.03,2 sets (DF=3) 3.04+/-0.08 and 3.14+/-0.04 and 1 set (DF=4) 3.00+/-0.05. 2) The actual DF values of 5 lV sets with an unmarked DF ranged from 3.04+/-0.05 and 4.80+/-0.06. 3) The percent deviation of the actual DF values from the theoretical values of DF= 1,3,4 and the unmarked DF sets were -6.5+/-3.0, 3.0+/-1.6, -24.1 and 0.7+/-15.7 (-23.6~20.1)% respectively. 4) Among the physical factors influencing flow rate the size of the inlet orifice size of the intra-venous infusion s('t was a major one. 5) The majority of intravenous infusion sets from various manufacturers in the ROK are unmarked with respect to DF or drops perml.


ABOUT
ARTICLE CATEGORY

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-3503, Lotte Castle President, 109 Mapo-daero, Mapo-gu, Seoul 04146, Korea
Tel: +82-2-792-5128    Fax: +82-2-792-4089    E-mail: journal@anesthesia.or.kr                

Copyright © 2024 by Korean Society of Anesthesiologists.

Developed in M2PI

Close layer
prev next